Electrical fast
±2kV for power-
transient/
supply lines
burst
±1kV for In-
IEC 61000-4-4
put/output
lines
±0.5, ±1kV line
Surge
to line
lEC 61000-4-5
±0.5, ±1, ±2kV
line to earth
<5 % UT
(>95% dip in
UT.)
Voltage
for 0.5 cycle
dips, short
<5 % UT
interruptions
(>95% dip in
and voltage
UT.)
variations on
for 1 cycle
power supply
70% UT
input lines
(30% dip in UT)
IEC 61000-4-
for 25 cycles
11
<5% UT
(>95 % dip in
UT)
for 250 cycles
Power fre-
quency (50/60
Hz)
30A/m
magnetic field
lEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
±2kV for power
supply lines
±0.5, ±1kV line to
line
±0.5, ±1, ±2kV
line to earth
<5 % UT
(>95% dip in UT.)
for 0.5 cycle
<5 % UT
(>95% dip in UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in UT)
for 250 cycles
30A/m
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the models
CanalPro X-Move requires
continued operation during
power mains interruptions,
it is recommended that the
models CanalPro X-Move be
powered from an uninter-
ruptible power supply or a
battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical lo-
cation in a typical commer-
cial or hospital environment.
33