11.1 System set-up

Damage to the attachment
caused by inappropriate tool!
CAUTION

11.1.1 Connecting the accessories

Risk of injury due to unap-
proved configuration using
additional components!
DANGER
Combinations of accessories that are not mentioned in
the present instructions for use may only be employed
if they are specifically intended for the respective
application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Please address your B. Braun/Aesculap partner or
Aesculap Technical Service with any inquiries in
this respect; for a contact address, see Chapter 15.
Use an appropriate tool for
changing the attachment
tool. Avoid contact with
cutting edges.
For all applied components,
ensure that their classifi-
cation matches that of the
application component
(e.g. Type BF or Type CF) of
the respective device.
11.1.2 Coupling and uncoupling of attach-
ments on the drill and reamer
Protection against inadvertent activation
Fig. 8
The rotational speed control button can be locked to
prevent inadvertent activation of the drill and reamer
during a tool change.
Note
For further information on the drill and reamer, see
TA011869.
Coupling
Push the attachment onto drill and reamer A until
it engages with a click, see Fig. 9.
Fig. 9
Uncoupling
Turn rotating sleeve B in the direction of the arrow
(clockwise when viewed from the rear end of drill
A