4.6
Using oxygen tubes (for Navy only
Oil
CO
CO H
O
2
2
2
1
1. Slowly open the valve for the compressed air supply.
2. Fit the T-piece (1) with 2 hose pieces, each approx. 5 cm
long and firmly insert into the tube holder H
3. Break off both tips on the oxygen tube (2).
4. Observe the green marking and arrow on the oxygen tube.
The arrow points in the direction of the accuro tube pump.
5. Connect the inlet side of the oxygen tube to the T-piece.
6. Connect the outlet side of the oxygen tube to the accuro
tube pump. Follow the applicable instructions for use.
7. Press the accuro tube pump (3) together as far as it will go
once, then loosen it and wait for the end of the stroke (white
indicator point clearly visible).
8. Close the valve of the compressed air supply. Immediately
remove the oxygen tube from the mount. Analyze the
results according to the instructions for use.
1
Optionally also with other devices.
1
)
3
O.
2
CAUTION
The oxygen tube heats up during the measurement. Do not
use in explosion-hazard areas, explosion hazard.
4.7
Using the MultiTest med. Int.
Inserting tubes (set includes adapter for air and O
The following media can be tested: O
Two signs can be found on the measuring device:
Sign 1
O
Air
2
Sign 1 specifies the media that can be tested. A color is shown
under the testable media.
Use
).
2
, air, CO
, N
O and N
2
2
2
CO
N
O
N
2
2
2
en
.
2
33