Information Messages - Storz ENDOMAT SELECT UP 210 Manual De Instrucciones

Ocultar thumbs Ver también para ENDOMAT SELECT UP 210:
Tabla de contenido
54
Technische Beschreibung
Brief description (1 line with approx. 40 characters possible)
Possible cause
102: Position the tubing set again
Last sensor test more than 24 h ago, deviation in pressure values
150: Pump stopped
Communication or pressure measurement disrupted
180: Paused
Transient discrepancy between redundant pressure measurement values
190: Cartridge detection error
Cartridge detection electronic error
191: Cartridge locking error
Cartridge locking electronic error
200: Hardware initialization error
Device defect
203: C2B error
Self-test: Communication error
20C: Restart necessary in < 4 h
The device has been in continuous operation for more than 20 h.
20D: Sensor test overdue
The device has been in continuous operation for more than 24 h.
251: C2B error
Internal communication error
252: Electronics error
Internal communication error
254: Default restored
File error
255: Electronics error
IO error
256: Date reset
RTC Battery empty
258: USB overcurrent
FE: USB power too high
259: Electronics error
BE: Internal error
Technical description
9. 4

Information messages

Response
Remove the tubing cartridge briefly and then start the pump again.
Device function interrupted, pump will need to be re-activated if error message is
automatically reset.
Transient sensor discrepancy, pump continues automatically
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
After 24 h of continuous operation, the pump can no longer be restarted.
Turn device off and on again.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Device function interrupted, pump will need to be re-activated if error message is
automatically reset.
Information message, work can continue, if necessary check the time setting
Information message, it may be possible to continue work
Information message, work can continue
Information message, work can continue, disconnect USB connection
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.