CS-3000 RF Generator Unit Operator Manual
RF ON/OFF Indicator (8)
The RF ON/OFF indicator lights up when RF energy is being transmitted. RF energy
transmission includes the periodic delivery of energy to measure impedance, even when the
RF generator has not been activated. However, when RF energy is being transmitted at the
power capable of causing coagulation, the RF ON/OFF indicator light is constantly
illuminated.
Connections
Front Panel Connectors
The Footswitch, RF Coagulation Cable for the Device, Indifferent, Dispersive Electrode and
Diagnostic Port for Sensing Cable CSK-2010 connectors allow interfacing with accessory
devices.
Depress the footswitch to begin delivering RF energy. To terminate the energy delivery and
reset the Time during operation, depress the footswitch again.
Indifferent, Dispersive Electrode (10)
The indifferent, dispersive electrode provides a path for the electrical current through the
patient and back to the generator. It is important to properly attach the indifferent, dispersive
electrode to the patient per manufacturer's instructions (see "Setup and Operation"). The
indifferent, dispersive electrode is for single use only.
Sensing Cable Connection (12)
The Sensing Cable Interface (Generator 'diagnostic' port) allows the connection of the
sensing electrodes from the EPi-Sense device to the external EP Sensing (EKG) equipment,
with the use of the CSK-2010 cable.
RF Coagulation Cable for the Device (9)
nContact Surgical manufactures the coagulation devices and RF coagulation cable for
exclusive use with the Model CS-3000 RF Generator Unit. Refer to the two Figures (5 & 6)
below, for the appropriate setup.
CSK Type Devices, CSK-2000 Cable and RF Generator Setup
Refer to the drawing below for the appropriate setup of the CSK devices with the CSK- 2000
cable and the CS-3000 generator.
Figure 5. CSK Type Devices, CSK-2000 Cable and RF Generator Setup
CDK Type Devices, CSK-2030 or CSK-2010, CSK-2000 Cable and RF
Generator Setup
Refer to the drawing below for the appropriate setup of the CDK devices with the CSK- 2000
cable, CSK-2030 or CSK-2010 cable and the CS-3000 generator.
Figure 6A. CDK Type Devices, CSK-2000, CSK-2030 Cable and RF
Generator Setup
Note: Pin Call out for CSK-2010 and CSK-2030 are the same
D1 = Distal Electrode 1 = Shrouded Pin #1; P1 = Proximal Electrode 1 = Shrouded Pin #3;
D2 = Distal Electrode 2 = Shrouded Pin #2; P2 = Proximal Electrode 2 = Shrouded Pin #4
nContact Surgical, Inc.
All manuals and user guides at all-guides.com
Footswitch (11)
Figure 6B. CDK Type Devices, CSK-2000, CSK-2010 Cable and RF
Generator Setup
Y
CAUTION:
Ensure that the EP Sensing (EKG) equipment comply with IEC 60601-2-
25 for protection from high frequency surgical interference.
Back Panel Connectors
Figure 7. Generator Back Panel – Key Features
Power Switch (13)
Switch that powers the generator on and off.
AC Power Connector (14)
Connector for the AC line power cable.
Grounding Stud (15)
Used as a ground equalization for safety and testing.
Serial Data Connector (16)
Serial communication connector to a host computer for data display and archival purposes.
Alarm Volume Control (17)
Knob for modifying the volume of the generator alarm. Rotate the knob clockwise to
increase the volume.
Chapter 2 Setup and Operation
Generator Setup and Operation
Preparing the Patient – Attaching the Indifferent, Dispersive Electrode
Prepare the patient for electrosurgery following standard protocol. Ensure patient's entire
body, including extremities, is insulated against contact with grounded metal parts. Closely
follow instructions for the coagulation device and manufacturer directions for the indifferent,
dispersive electrode.
Y
CAUTION: Failure to achieve good skin contact by the entire adhesive surface of
o
the indifferent, dispersive electrode could result in a patient burn or poor electrical
performance from the coagulation device. Note: Patient Return Electrode (e.g.
Indifferent, Dispersive Electrode), surface area of 21 square inches (136cm
equivalent to Covidien Valleylab Polyhesive Reference Number E7506 is
recommended.
Setting up the CS-3000 RF Generator
1.
Connect the supplied power cord into back mains receptacle of the CS-3000 RF
Generator.
2.
Plug the CS-3000 RF Generator power cord into an outlet.
3.
On the back of the CS-3000, activate the mains switch and turn the RF Generator on.
4.
Turning on the generator (switch on rear of generator) causes the system to enter
Standby mode where no measurements or settings are possible. The nContact logo
and the software version number are displayed in the message window; the Standby
LED is illuminated.
LBL-2123 Rev. E