CS-3000 RF Generator Unit Operator Manual
Chapter 6 Troubleshooting
The following flowchart illustrates a user decision tree to assist in operation of the
generator and troubleshooting.
Figure 19. Generator Modes, States and Operational Flow
WARNING: Do not open the back panel of the RF Generator. This may cause serious
injury and damage to the unit. It will void the warranty. When problems cannot be
resolved by the directions in this troubleshooting section, contact nContact Surgical, Inc.
for additional service and repair information
Table 1 provides a list of symptoms which may occur during routine operation of the RF
Generator. If you encounter a problem that is not listed here, contact nContact Surgical,
.
Inc
Table 1 – List of Troubleshooting Symptoms & Actions
Symptom
No displays or indicators when the
RF Generator is turned on
IMP LIMIT warning, indicating
impedance out of range
nContact Surgical, Inc.
All manuals and user guides at all-guides.com
Action
• Be sure the generator is plugged into a
working electrical outlet.
• Check power switch on back of generator.
• Unplug and check the fuse on the rear
panel. Fuses should only be replaced with
LittleFuse 2183.15, Time Delay (Slo Blo)
rated at 3.15A, 250VAC.
• Check connections to coagulation device
and indifferent, dispersive electrode.
• Ensure device is properly placed on tissue
site.
• Check position of indifferent, dispersive
electrode on patient's back or thigh.
• If problem persists, replace the coagulation
device and indifferent, dispersive
electrode.
Symptom
Error LED indicator illuminates and
one of the following Fault
messages appears in Message
window:
RF EFF LIMIT
ROM CRC FAILURE
RAM FAILURE
TIMER FAILURE
+48 VOLT SUPPLY FAIL
-12 VOLT SUPPLY FAIL
+12 VOLT SUPPLY FAIL
+5 VOLT SUPPLY FAIL
IMPEDANCE TEST FAIL
RF POWER TOO HIGH
WHEN OFF
RF POWER TOO HIGH
WHEN ON
RESISTIVITY FAILED
MEASURED POWER
CALCULATED POWER
FAN FAULT
GROUND FLOAT ON A/D
CAL VOLTAGE ON A/D
AMBIENT TEMP FAIL
Error LED indicator shows a button
is stuck and one of the following
Fault messages appears:
RF ON BUTTON STUCK
MODE BUTTON STUCK
POWER/TEMP UP
BUTTON STUCK
POWER/TEMP DOWN
BUTTON STUCK
TIME UP BUTTON STUCK
TIME DOWN BUTTON
STUCK
RF power does not turn on when
footswitch is depressed
RF interferes with ultrasound and
other equipment
Impedance is greater than 500 Ω at
the beginning of RF power delivery
Impedance is less than 30 Ω at the
beginning of RF power delivery
LBL-2123 Rev. E
Action
• Power unit OFF, then Power ON again.
• Allow generator to run through normal
start-up self-diagnostics.
• If generator returns to Error state and the
problem persists, contact your nContact
Surgical, Inc. representative.
• Depress and release the indicated button
to determine if the button remains stuck.
• Power unit OFF, then Power ON again.
• Allow generator to run through normal
start-up self-diagnostics.
• If generator returns to Error state and the
problem persists, contact your nContact
Surgical, Inc. representative.
• Verify the footswitch is connected to the
front panel of the generator
• Ensure Time is not set to 0.
• Ensure IMP LIMIT error is not detected -
impedance should be within 30 – 500Ω.
• Ensure a coagulation device is connected
to the generator
• Ensure the indifferent, dispersive electrode
is attached to the patient and connected to
the generator
• Check the footswitch by unplugging its
cable from the generator, placing a thumb
over the foot pedal connector then
depressing the pedal. If the footswitch is
functioning properly, air should be expelled
through the connector as the footswitch is
depressed.
• Ensure the cables from the electrodes do
not cross the cables from the ultrasound
probe or other equipment.
• Changing settings on the ultrasound may
alleviate image interference.
• Ensure the indifferent, dispersive electrode
is properly attached.
• Check all connections.
• Clean off any coagulum from the
coagulation device.
• Ensure device is properly placed on tissue
site.
o
Reposition the coagulating device
Page 12/ 79