CanalPro Jeni
Sterilisation and storage
Sterilisation should be carried out in compliance with EN 13060, class B
Please observe the instructions of the device manufacturer.
Prior to sterilisation, the contra-angle has to be cleaned, disinfected and
lubricated with oil.
For sterilisation, please make sure to seal the contra-angle in a suitable
sterilisation packaging for sterile articles in compliance with EN 868-5.
Please make sure that the device to be sterilised is absolutely dry.
Store the sterile device in a dust-free and dry place.
Approved sterilisation methods
Please observe all specific guidelines, standards and provisions in your country
Steam sterilisation class B with sterilisers as per EN 13060.
Minimum sterilisation holding time 3 minutes at 134 °C or
Steam sterilisation class S with sterilisers as per EN 13060
Must be expressly authorised by the manufacturer of the sterilizer for
sterilisation of contra-angles.
Minimum sterilisation holding time 3 minutes at 134 °C
Accessories
The equipment used must be suitable for dental instruments and contra-angles.
For example:
W&H Service Oil F1, MD-400 and ISO-E spray nozzle with spray adapter for
handpieces and contra-angles
22