CanalPro Jeni
Lubrication
1. Lubricate the contra-angle with oil,
for example with W&H Service Oil
F1, M400. Please follow the
instructions of use printed on the
aerosol can.
or
2. Lubricate the contra-angle in an
automatic instrument maintenance
device, for example in the W&H
Assistina (see instructions for use coming with the device). Recommended
maintenance cycles: Imperative after each internal cleaning, before each
sterilization, or after 30 minutes of use, or once a day. After maintenance, please
place the contra-angle in an upright position on a clean disposable cloth. This way,
excessive oil may drip off so that it cannot get into the motor.
Testing
Lubrication
1. Hold the contra-angle with the head
pointing downwards.
2. Let the contra-angle run briefly to
remove excessive oil. To this end,
start the calibration function on the
controller. In case of dirt leaking from
the contra-angle, the entire hygiene
and maintenance procedure must be
repeated.
3. Wipe the contra-angle with absorbent paper or a soft cloth.
after
21