1.3 EN/IEC 60601-1-2:2007 Sub-clause 5.2.2.2 Table 4
The Coltolux
LED unit is intended for use in the electromagnetic environment specified below . The customer or end user of
®
the Coltolux
LED unit should assure it is used only in such an environment .
®
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80MHz and 8000MHz, the higher frequency range applies .
NOTE 2 These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorbtion and
reflection from objects, structures and people .
Field charge from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land mobile radios,
amature radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy . To access the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered . If the measured field stren-
gth in which the location of the Coltolux
LED unit should be observed to verify normal operation . If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or re-locating the Coltolux
Over the frequency range 150kHz to 80 Mhz, field strengths should be less than 3V/m .
Guidance and manufacturers Declaration - Electromagnetic Immunity
IEC 60601 Test Level
3V rms 150kHz to 80 MHz
3V/m\
80MHz to 2 .5 GHz
LED unit is used exceeds the application RF compliance level above, the Coltolux
®
Compliance Level
3V rms 150kHz to 80 MHz
3V/m\
80MHz to 2 .5 GHz
LED unit .
®
Intended Electromagnetic Environment
Portable and RF communications
equipment should be used no closer
to any part of the Coltolux
LED unit,
®
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter .
Recommended separation distance
d = 1 .2√P
d = 1 .2√P 80MHz to 8000Mhz
d = 1 .2√P 800Mz to 2 .5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufactu-
rer and d is the recommended minimum
separation distance in meters (m) .
Field strengths from fixed RF transmit-
ters, as determined by an electroma-
gnetic site survey, should be less than
the compliance level in each frequency
range .
Interference may occur in the vicinity of
equipment marked with the following
symbol:
9
®