Select Main Menu to return to home screen
At anytime the Smart Monitor screen is in sleep mode or powered down, to continue push the
Home Button and enter "000000" as the passcode to access the Smart Monitor.
New Case
c.
Ensure the power supply is connected to mains power and power-on Smart Monitor. The Smart
Monitor will function without being connected to the mains power, but it is not recommended.
1) Assemble disposables to applicable Cath lab components (e.g. manifold) and perform
priming per standard Cath lab procedure and the disposable Instructions for Use.
2) From the Main Menu select New Case.
If Case ID and Renal Function was enabled in User Settings then enter information
desired to be stored in the case database, select "Next" and confirm the information
entered is correct. Once the data entered is confirmed the Smart Monitor will scan for
available disposable devices to connect with.
If Case ID and Renal Function was disabled in User Settings then the Smart Monitor will
scan for available disposable devices to connect to.
3) Remove battery pull tabs from disposables.
Note: It is recommended to leave all battery pull tabs in the disposable devices until all
priming is complete and the case is ready to begin. This will ensure maximum battery life
and performance throughout the case.
4) Confirm wireless connection of the Smart Monitor with the selected disposables.
Following scanning, the LEDs on the disposables will flash in the same pattern as shown
on the Smart Monitor for proper identification. ENSURE that the flashing LED pattern on
the disposable matches the corresponding flashing pattern on the Smart Monitor. Failure
to confirm pattern could allow the Smart Monitor to connect to disposables intended to be
connected to a different Smart Monitor.
Select Yes on the Smart Monitor to confirm Syringe flashing pattern.
Select Yes on the Smart Monitor to confirm Module flashing pattern.
Note: Selecting No will result in further scanning.
Note: When the device is connected to the Smart Monitor, the green LED on the
disposable device will remain on for 10 seconds.
Alternate Wireless Connection Confirmation
Confirmation of connection of the Smart Monitor to the module disposables can be accomplished
by pressing the pause button of the disposable if available; and confirmation of connection to
the Smart Syringe can be accomplished by moving the plunger > 2 mL.
5) Enter Physician Specified Threshold Volume: Enter value (mL) for physician specified
threshold volume or "0" for no threshold.
DO NOT select Start Case until device priming is completed.
6) Start Case
ENSURE ALL PRIMING IS COMPLETED BEFORE PRESSING Start Case.
Recommend to not press Start Case until just prior to first injection of contrast to the
patient.
Select Start Case to begin contrast accounting.
Perform procedure.
Pause/Resume System Operation
d.
When system is in "Pause" mode, cumulative contrast accounting is suspended. "Resume"
mode allows contrast accounting.
To Pause
Pause
Select Pause on Smart Monitor, turn
Lock
DyeVert Stopcock "OFF" to contrast
Disabled
savings (diversion) or press DyeVert
Plus Module pause button. LED's on
Smart Syringe & Module will blink
yellow and background on Smart
Monitor will change to yellow
Pause
Select Pause on Smart Monitor or
Lock
press DyeVert Plus Module pause
Enabled
button. LED's on Smart Syringe &
Module
will
blink
yellow
background on Smart Monitor will
change to yellow
If contrast is injected back to the source, the system does NOT need to be paused. The system
will automatically recognize this as not being administered to the patient.
System Pause:
select Pause OR press Pause Button
To Resume
Aspirate contrast into Smart Syringe
(resumes
automatically),
turn
DyeVert Plus Stopcock "ON" to
contrast savings (diversion), select
Resume on Smart Monitor, or press
DyeVert Plus Module pause button.
LED's on Smart Syringe and Module
will be green for 10 seconds and
Smart Monitor will return to normal
background
Select Resume on Smart Monitor, or
press DyeVert Plus Module pause
button. LED's on Smart Syringe and
and
Module will be green for 10 seconds
and Smart Monitor will return to
normal background.
8280-B