Guidelines and manufacturer's statement - electromagnetic compatibility
EN
42
14.6
Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity - for the Device FLOSTEADY Arthroscopy
Pump
Electromagnetic
Test level
interference
immunity tests
Conducted HF
3 V
eff
interference
150 kHz to
quantities
80 MHz
according to IEC
61000-4-6
3 V/m
Radiated HF
80 MHz to
interference
2.5 GHz
quantities
according to IEC
61000-4-3
Note 1: The higher frequency range applies for 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
Compliance
Electromagnetic environ-
ment guidelines
In compliance
Portable and mobile wire-
less devices should not be
used in closer proximity to
the device FLOSTEADY
Arthroscopy Pump (includ-
ing cables/lines) than the
In compliance
recommended safety dis-
tance calculated based on
the transmitting frequency
and the applicable for-
mula. Recommended
safety distance:
d = 1.2P for 150 KHz
to 80 MHz
d = 1.2P for 80 MHz to
800 MHz
d = 2.3P for 800 MHz
to 2.5 GHz
With P as the rated output
of the transmitter in watts
[W] according to the infor-
mation provided by the
manufacturer of the trans-
mitter and d as recom-
mended safety distance in
meters [m].
The field strength of sta-
tionary transmitters for all
frequencies tested on site
a
should be lower than the
concordance level.
Interference is possible in
the proximity of devices
featuring the following
pictograph.
b