Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains connection cable may
be plugged only into a properly installed, grounded safety wall socket (shock-
proof socket) (see DIN VDE 0100-710). Read the device label located in rear of de-
vice (type plate) to determine the operating voltage of the device.
WARNING!
To avoid the risk of electrical shock, only use this device when connected to a
properly grounded power supply network.
Integrate the device into the potential equalization system as specified by local
safety rules and regulations.
Medical devices are subject to special safety and protective measures concerning
electromagnetic compatibility (hereafter abbreviated as EMC).
This device is to be used only for the purposes described in the manual and has
to be installed, set up, and operated in compliance with the EMC notes and in-
structions.
Initial Device Startup
Potential equalization
Precautionary measures