Technical Data
EN
Mains connection cable:
Frequency:
Max. Power consumption:
Max. Current consumption:
Classification according to Directive 93/42/EEC:
Protection class:
Protection level:
Moisture protection:
Dimensions:
without toggle screw
incl. toggle screw
Weight:
Operating conditions:
Storage and transportation conditions:
Manufactured and tested according to:
EMC:
Adjustable pressure range:
Overpressure safeguard:
Capacity/output:
38
13
Technical Data
100-240 V~
2x T3, 15 A, UL-recognized
Connection for potential equalization
50-60 Hz
Mean value: 70 VA
Peak value: 140 VA
100 V: 1.2 A
240 V: 0.6 A
IIa
II
Type BF
IP42
Width x Height x Depth
145 x 215 x 170 [mm] / 5.71 x 8.46 x 6.69 [inch]
145 x 215 x 230 [mm] / 5.71 x 8.46 x 9.06 [inch]
approx. 4.7 kg / 10.3 lbs
10-40°C /50-104°F
30-75% rel. humidity
700 - 1060 hPa ambient pressure
-30 - +70° C /-22 - +158° F
10-85% rel. humidity
85 to 100% rel. humidity
EN 60601-1 / IEC 60601-1
IEC 60601-1:1998 +A1:1991+A2:1995
IEC 60601-1:2005+Corr2006+Corr2007+Int1:2008+Int2:2009
EN 60601-1-2 / IEC 60601-1-2
15-200 mm Hg
Motor Cutout at 250 mm Hg
0,1 - 3.0 l/min