Pdx Cable Preparation - Panel - Mettler Toledo IND560 PDX Guia De Instalacion

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

IND560 PDX Installation Manual

PDX Cable Preparation - Panel

Ground and shield terminations are a critical part of the PDX system's immunity to
noise and electrical surges. Prepare and install the terminal end of the PDX home
run cable as follows.
If the panel-mount IND560 PDX terminal is being installed inside an enclosure
and the PDX cable will enter through a gland, prepare the cable in the same
way cables are prepared for installation into a harsh IND560 PDX. This is
shown in Figure 2-18. Extend the length of the ground wire (Figure 2-19) and
the red, white and blue wires so they are long enough to reach the rear of the
IND560 PDX. The internal grounding and shielding do not need to be
extended, as they will terminate at the gland where the cables enter the
enclosure; install them as for the harsh enclosure.
Figure 2-25: Internal Grounding and Shielding, Enclosed Panel Mount Terminal
If the IND560 PDX will be installed into a control cabinet where an entry gland is
not used, prepare the cable per the following:
1. Cut and remove an appropriate length of the outer braided shield, the
insulation sheath, the inner braided shield and insulation as shown in Figure
2-26. The suggested lengths are listed in Table 2-4.
2-16
PDX cabling
Cable sheath
C
A
B
Figure 2-26: PDX Cable Preparation – Panel
Enclosure
Ground wire
extension
Enclosure
grounding stud
Cable gland
Ground wire
D