5. The hardware to mount the terminal to the vertical surface is not included with
the terminal—it must be supplied locally. Ensure that the mounting hardware
is capable of supporting the weight of the terminal, which is approximately 3.5
kg (8 lb). Using the locally supplied hardware, mount the terminal to the
vertical surface.
Installing Cables and Connectors
Information for installing cables and connectors for the IND560 PDX terminal is
provided in this section, including:
•
Ferrites
•
Harsh Enclosure Cable Glands
•
Main Board Wiring Connections
•
Wiring Connections for Options
Ferrites
In order to meet certain electrical noise emission limits and to protect the IND560
PDX from external influences, it is necessary to install a ferrite core on some of the
cables connected to the terminal. There are two ferrite cores included with the basic
terminal and additional ferrites are supplied with each of the options.
To install ferrites, simply route the cable through the center of the core and then
take one wrap around the outside of the core and route the cable through again.
Either the complete cable or the individual wires can be wrapped through the
ferrite. This should be done as close to the enclosure as possible. See Figure 2-12.
IND560 PDX Installation Manual
Figure 2-11: Mounting Hole Pattern
2-7