10. Attach the ground lug (from the platform's lightning protection kit) to the rear of
the enclosure using one of the two screws provided. Refer to Figure 2-23.
Outer shield
drain wire
Flat braided
11. Route the home run cable's outer shield drain wire and the flat braided cable
through the ground lug attached to the rear of the enclosure. Figure 2-23
shows the cables clamped in position.
12. The remaining end of the flat braided cable is terminated at the ground rod as
indicated in the scale platform's installation drawing.
13. Inside the enclosure, connect both the single short black wire (from Figure
2-19) and the twisted black and green wires from the PDX cable to the internal
ground stud with a hex nut as shown in Figure 2-24.
2-14
Clamp
Figure 2-22: Securing the Outer Shield – Harsh
cable
Figure 2-23: Terminal Chassis Grounding Connections – Harsh
Outer shield
drain wire
Flat braided
ground cable
Outer braided
shield
External
ground lug
Remaining end
of flat braided
cable