IND560 PDX Installation Manual
Power Requirements
The terminal requires 85 to 264 VAC (at 750 mA maximum) with a line frequency
of 49 to 61 Hz of power and is internally fused at 1.6 amp, 250 volts. The fuse is
located adjoining the main power connection on the main board as shown in
Figure 2-16. Should the fuse blow, always replace the fuse with the same voltage
and amperage as specified and have a qualified METTLER TOLEDO service
provider test the power supply for correct operation.

PDX Cable Preparation - Harsh

Ground and shield terminations are a critical part of the PDX system's immunity to
noise and electrical surges. Prepare and install the terminal end of the PDX home
run cable as follows:
1. Cut and remove an appropriate length of the outer braided shield, the
insulation sheath, the inner braided shield and insulation as shown in Figure
2-18. The suggested lengths are listed in Table 2-3.
Letter
2. Prepare a separate black ground wire as shown in Figure 2-19 using the
same gauge black wire as in the PDX cable. Crimp one of the two included
loop terminals onto the end of the wire as shown.
2-12
E
B
C
A
Figure 2-18: PDX Cable Preparation - Harsh
Table 2-3: Cable Preparation Lengths - Harsh
Description
A
Outer braided shield to end of cable sheath
B
Inner braided shield and foil
C
Black and green wires and internal drain wire
D
Red, white and blue wires
E
External shield drain wire
Figure 2-19: Black Ground Wire - Harsh
D
Length
30 mm (1.2 in)
8 mm (0.3 in)
50 mm (2.0 in)
210 mm (8.25 in)
100 mm (4.0 in)