Figure 2-53: PROFIBUS Option Board, Panel Mount Enclosure
Pin assignments for the PROFIBUS interface are shown in Figure 2-54.
PIN
Figure 2-54: PROFIBUS Nine-Pin Connector Assignments
Follow the wiring instructions included with the connector to terminate the wires.
EtherNet/IP and Modbus TCP Interface
The EtherNet/IP Module (Figure 2-55) connects to the network via a standard
Ethernet patch cable. The module's address is set in software, and the DIP
switches are not used and must all be set to OFF.
Figure 2-55: Ethernet / IP Option Board
IND560 PDX Installation Manual
NOTES:
SIGNAL
1. USE MATING CONNECTORS AND CABLE
1
Not used
RECOMMENDED FOR PROFIBUS CONNECTIONS.
2. REFER TO PROFIBUS INTERNATIONAL
2
Not used
DOCUMENTATION FOR OTHER CONSIDERATIONS.
3
RxD/TxD +
4
RTS
5
GND bus
6
+5V bus
7
Not used
8
RxD/TxD -
9
Not used
2-33