Protective Suit with Integral Full Face Mask P or with Inte-
grated Full Face Mask PE or with Integrated Full Face Mask
PE/ESA
Additional test equipment:
Adapter – R 53 345
Exhalation valve test plug – R 53 346
Remove protective cap from
connector.
Press spring carrier inwards at
one end and remove it.
Grasp valve disc at nipple and
remove from guide.
Insert the exhalation valve test
plug and fix it with the spring car-
rier. Insert spring carrier so that
both hooks snap into the slots at
the side. The spring carrier is
marked "L" = left and "R" = right
and is positioned at an angle.
Check adapter; the bore must be
open end-to-end. Widen the bore
with a 5 mm bit if necessary.
Lightly press adapter into connec-
tor.
1 Fit the test cap and connect the
air gun.
®
2 Connect Porta Control
and adapter.
Inflate chemical protective suit
and test for leaks (see "Checking
the Chemical Protective Suit for
Leaks" on page 26).
Checking the Chemical Protective Suit for Leaks
To avoid any overloading of the Porta Control
ure alternately:
1 Open hose clip for air supply and briefly actuate air gun,
then reclose hose clip.
2 Open the tube clamp for Porta Control
sure at the pressure gauge, then close the tube clamp.
Inflate the protective suit until it is free of folds and Porta
®
Control
displays 17.5 mbar.
Close hose clips.
Set the test time to 10 minutes and start the stop watch.
Hold pressure at approx. 17 mbar during this time so that
temperature and pressure can equalise. Boost pressure via
hose clip and air gun if necessary.
26
Disconnect the compressed air gun and open the tube
clamp. Lower the pressure to 16.5 mbar and close the tube
clamp. Set the test time to 6 minutes and start the stop
watch. Read off the pressure on Porta Control
test time has run down.
If the pressure drop is smaller than or equal to 3 mbar, the
protective suit is considered as tight. In this case, disassemble
the test set-up and check the suit valve.
If the pressure drops by more than 3 mbar: coat critical areas
such as seams, cuff, zip fastener, glove and boot connections
with soapy water and mark any leaks, then rinse off soap solu-
tion, ventilate and repair suit. Then repeat leak test. Alterna-
tively, the chemical protective suit can be sent to Dräger for
repair.
Testing the Suit Valve
Test equipment:
Test cap – R 53 289
Test plug – R 53 287
Moisten valve disc and insert it.
Open the zip fastener.
Press test plug into suit valve from
inside.
Assemble test set-up. Note direc-
tion of arrow on pumping ball.
1 Open hose clip and create a neg-
ative pressure of 10 mbar with
pumping ball, then reclose hose
clip. Do not overload pressure
gauge.
2 Set the test time to 1 minute and
start the stop watch.
Read off pressure at end of test
time.
If the change of pressure is less
than 1 mbar: The suit valve is OK. Disassemble test set-up
2
and refit protective cap. The opening of the protective cap
1
points towards the feet.
If the change of pressure is greater than 1 mbar:
Dismount and visually check the valve disc.
Valve disc and seat must be clean and undamaged. Replace if
necessary. Moisten valve disc, insert and repeat test.
®
, fill and meas-
After Testing
Pull the test plug out of the suit valve and install the protec-
®
tion cap.
, read off the pres-
Keep a record of the testing (see "Record Card" on
page 36).