Select Menu > User Maintain > Network > Upload Setting > Config and set Delete Data After Upload
to On, the successfully uploaded data will be deleted from the device.
Alarms
Alarm levels
The sound pressure range for standard audible alarm signals is from 45 dB to 85 dB, with the
measurement radius of 1 meter.
Nurse Call
The monitor provides dedicated nurse call port which is connected to nurse call system through the nurse
call cable to perform the nurse call function. Enter Menu > User Maintain > Common > Nurse Call to
choose Off, High, Med., Low, High+Med. or All (default). When the corresponding level of physiological
alarm occurs or when the nurse call cable falls off, the nurse call signal will be sent to remind the nurse.
Select the shortcut key
NOTE
Before using the function of nurse call, check if it is functioning normally.
NOTE
Nurse call function will be inactivated when the monitor is in audio alarm paused/off status.
Monitoring SpO
Measurement limitations
Inaccurate measurements can be caused but not limited by:
darkly pigmented skin
Assessing the validity of a SpO
NOTE
The SpO
accuracy has been validated in controlled human studies against arterial blood sample
2
reference measured with a CO-oximeter. SpO
two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-
oximeter measurements. The volunteer population in the studies are composed of healthy men and
women from 18 to 50, with various skin pigmentations. Note that the study population was healthy adults
and not in the actual intended use population.
8
to reset the alarm.
2
reading
2
measurements are statistically distributed, only about
2
Vista 120 SC – Supplement