Dräger Vista 120 SC Suplemento Del Manual página 3

Ocultar thumbs Ver también para Vista 120 SC:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 105
English
Supplement to the Instructions for Use
Vista 120 SC Instructions for Use Part no. 2680025
This supplement is used for Vista 120 SC SW 1.n.
 This supplement must be kept with the Instructions for Use of the medical device.
 This supplement updates the information given in the following chapters of the Instructions for Use.
For your safety and that of your patients
Product-specific safety information
WARNING
In case of power interruption, if the power is restored, monitoring will resume with all active settings
unchanged.
WARNING
Always remove non-defibrillation-proof accessories during defibrillation.
WARNING
If the accessory is accidentally disconnected, it shall be reconnected by a trained healthcare
professional.
CAUTION
Ensure that the monitor is supplied with continuous electric power during operation from the main source
power or batteries. Sudden power failure may cause failure to monitor parameters.
NOTE
The materials with which the patient or any other person can come into contact conform to the standard
of EN ISO 10993-1.
Remove the description below:
The simultaneous use of cardiac pacemaker and other patient-connected equipment may cause safety
hazard.
Remove a battery whose life cycle has expired from the monitor immediately.
In this Instructions for Use, "MRN" is changed to "Patient ID", "Bed No." is changed to "Bed Label", "Vista
120 CMS" is changed to "Vista CMS", "Vista 120 Gateway" is changed to "Vista Gateway", "Zero
Calibration" is changed to "Zero", and blue alarm indicator is changed to cyan.
Vista 120 SC – Supplement
Supplement
Vista 120
en
3
loading