Maintenance
Inspection
The overall check of the monitor, including the safety check, should be performed only by qualified
personnel every 24 months, and each time after fix up.
The following items should be checked:
 If the environment condition and power supply meet requirement.
 If the power supply cord has damage and insulation meets requirement.
 If the device and accessories have damage.
 Specified accessories.
 If the alarm system can work properly.
 If the recorder can work properly and the paper meets the requirement.
 Battery performance
 If all functions are in good conditions.
 If the grounding resistance and leakage current meet requirement.
If any damage or abnormality is found, don't use the monitor and contact local Customer Service Center.
Maintenance shall be carried out at least once every two years, or as specified by local regulations. The
following tasks are for Dräger-qualified service professionals only. Contact an Dräger-qualified service
provider if the monitor needs a safety or performance test. Clean and disinfect equipment to
decontaminate it before testing or maintaining it.
Perform inspections at regular intervals and observe the following specifications.
Checks
Inspection/safety checks
Metrological checks
Dräger will make available on request circuit diagrams, component part lists, descriptions, calibration
instructions, or other information that will assist service personnel to repair those parts of the equipment
that are designated by Dräger as repairable by service personnel.
Metrological checks
If required by applicable regulations, the following measurement functions must be checked every two
years by qualified DrägerService personnel.
 Body temperature
 Non-invasive blood pressure
16
Interval
Every 2 years
Every 2 years
Personnel Responsble
Service personnel
Service personnel