Networked monitoring
Connecting the monitor to Vista CMS
WARNING
Due to the delay of network transmission, data viewed at the Vista CMS has a delay of less than
10 seconds compared with data generated at the corresponding monitoring device. To avoid potential
patient injury, do not rely exclusively on the system when it comes to time-critical observations.
NOTE
The delay time from onset of ALARM CONDITION to the point where representation of ALARM
CONDITION leaves the SIGNAL OUTPUT PORT is less than 1 second.
NOTE
The sum of the mean alarm condition delay plus the mean alarm signal generation delay is less than 10 s.
Connecting the wireless network
NOTE
It is not recommended to use Wi-Fi and Bluetooth simultaneously; otherwise, co-channel interference
may be arisen.
Inquire for patient information via network server (ADT)
NOTE
Patient information (other than patient type) obtained from network server cannot be edited.
Uploading data to network server
Ward round records and spot-checking data can be uploaded to network server.
Upload manually: Click Upload in the Review interface to upload the stored ward round record or the real-
time spot-checking measurement data.
Upload automatically: In User Maintain > Network > Upload Setting > Config > Real-time Upload, if it
is set to On, when monitor is in networking and is also connected with network server, the monitor will
upload the stored ward round record or the real-time spot-checking measurement data to network server
in real time.
Set the Timeout Duration as needed in User Maintain > Network > Upload Setting > Config. If the data
is not uploaded within the predefined time, the monitor will prompt upload failure. The upload result will be
prompted in the main interface, and in review interface, the icon
successfully sent to the network server.
Select Menu > User Maintain > Network > Upload Setting > Config and enable Backfill Resume.
When the monitor is in idle and there are data failed to be uploaded, the monitor can automatically upload
those data from the breakpoint if the network server is connected.
Vista 120 SC – Supplement