Check whether a glass tube fuse is defective.
Replace a defective glass tube fuse with a
functioning fuse of the same type.
Insert the fuse holder together with the glass tube
fuses.
Lubricating
The calibration and adjustment facility is equipped
with a central lubrication system. The lubricating
nipple (1) is mounted behind a cover on the front
right-hand side.
We suggest Fin Grease MP 2/3 from Interflon.
The calibration and adjustment facility must be
lubricated every three months.
Remove the cover of the lubricating nipple.
Lubricate the lubricating nipple with two pumps of
the grease gun.
Dispose of escaped grease and consumable
materials soiled with grease according to the
regulations applicable at the operating location.