Stahlwille perfectControl 7794-2 Traduccion Del Manual De Instrucciones Originale página 82

Dispositivo de ajuste y calibración con motor
Ocultar thumbs Ver también para perfectControl 7794-2:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 150
Calibrating and adjusting
torque wrenches
CAUTION
Risk of injury due to the torque wrench
breaking. Parts can fly away.
Make sure that the protective cover
is in flawless condition.
Do not operate the device with a
damaged protective cover.
Exchange a damaged protective
cover for an undamaged one.
Only operate the calibration and adjustment
facility if all protective facilities are complete and
functional.
Make sure that nobody can be endangered by the
calibration and adjustment facility's starting up
before switching it on.
Starting the calibration and adjustment
facility
Switch the calibration and adjustment facility on
using the switch on the IEC power connector.
Position the transducer on the flange using the
quick-release system.
Connect the transducer to the calibration and
adjustment facility using the helical cable.
To exchange the transducer, press the two green
locking pins and raise the transducer.
Do not apply force when doing this.
Calibrating and adjusting torque wrenches
Set-up mode
After switching on or after pressing the stop button
at the end of a calibration process, the calibration
and adjustment facility is in set-up mode.
The yellow LED in the start button lights up.
To turn the transducer counter-clockwise, press
the start button.
To turn the transducer clockwise, press the stop
button.
Move the transducer to the neutral position by
pressing the start or stop button. Align the two
marks to do this.
33