6
Configurable Default Rhythm Speed
CP 200
You can now change the default rhythm speed: 10, 25, or 50 mm/s.
To change this default, press the Menu key
Rhythm Report Settings > Default Rhythm Speed.
Transferable Configurations
CP 100, CP 200
You can now transfer the configuration of one electrocardiograph to another (CP 200 to
CP 200, or CP 100 to CP 100). If you have the spirometry option, its configuration is
included in the transfer.
To transfer a configuration, follow these steps:
1.
Insert a storage device (SD memory card or USB flash drive) into an
electrocardiograph that is configured as desired.
2. Press the Menu key
3. Choose System Settings > Device Info > Manage Settings > Export Files >
Export Configuration Files.
4. The files are copied to the storage device.
5. Remove the storage device, and insert it into another electrocardiograph of the
same model.
6. Choose System Settings > Device Info > Manage Settings > Import Files >
Import Configuration Files.
7 .
The files are copied to the electrocardiograph, and then it reboots.
CE Mark Change
Welch Allyn has changed its CE Mark to
. Choose ECG Settings >
.
Configuration file
or
.
Welch Allyn CP Series Electrocardiographs