Chapter 5 Performing ECG Tests
Figure 33. Auto ECG Acquisition" Screen
Doe, Jane
Doe, Jane
Auto ECG : Acquisition
Auto ECG Acquisition
Collecting Data
Gain: 10 mm/mV
Speed: 25 mm/s
Baseline Filter: ON
Muscle Filter: ON
4. If prompted, choose whether to wait for the electrocardiograph to acquire 10 seconds
of filtered, processed data before printing.
If you override the wait time and print the available data immediately, be aware that
the printed data will be insufficient in quality or quantity or both.
The report prints.
Note
After printing, the Auto ECG Post-Print screen appears.
9:17AM Oct 16 05
9:17AM Oct 16 05
Printing Report
If you do not enter the patient data, write the patient identification information on
the printed report immediately after the ECG test to avoid associating the ECG
report with the wrong patient.
ECG report Print Copy feature will not be available if no Patient Name or ID are
entered.
If a red stripe appears along the edge of your report, replace the paper. See
"Loading the Thermal Chart Paper"
Welch Allyn CP 100 Electrocardiograph
HR 120
The electrocardiograph displays messages here
when it is doing the following:
• Collecting data
• Analyzing data
• Printing the report
on page 21.
Figure 34
on page 51.