Compliance With 60601-2-51
CP 100, CP 200
The electrocardiograph is in compliance with regulatory standards CAN/CSA C22.2
No. 60601.2.51 and IEC/EN 60601-2-51. For manufacturer's disclosures regarding
60601-2-51, please contact Welch Allyn Customer Service at the phone number listed in
the Directions for Use.
Pacemaker Detection Prompt
CP 100, CP 200
If the software detects the possible presence of a pacemaker, it now asks you whether
the patient has a pacemaker. If you say no, interpretation (if purchased) is included in the
report. If you say yes, interpretation is not included in the report. This prompt was added
to prevent false pacemaker pulses caused by environmental interference.
Symbols and International Characters
CP 100, CP 200
You can now enter symbols and international characters using the keyboard.
To enter a symbol, hold the Shift key while pressing a number key. To enter an
international character, hold the Ctrl key while pressing a key with an international
character above it on the keyboard label. To capitalize an international character, hold
Shift + Ctrl while pressing the key.
CAPS Lock
CP 100, CP 200
You can now turn CAPS Lock on and off in the software. This feature works just like a
standard Caps Lock key.
To turn CAPS Lock on or off, press the Menu key
Choose System Settings > Device Configuration > CAPS Lock > On or Off.
Swedish, Danish, and Finnish Patient ID Processing
CP 100, CP 200
When you enter a valid Swedish, Danish, or Finnish patient ID, the patient's date of birth/
age, gender, and race are now automatically filled in.
"Repeat Test" Softkey
CP 200
The ECG post-print screen now includes a Repeat Test softkey to let you more efficiently
repeat a test.