INSTRUCTION MANUAL
PARAMETER
FACTORY
80L
150L
A
120 120
B
15
20
D
3
3
E
180 180
F
30
40
M
120 120
N
+1.05 +1.05
U
85
85
S
1.2
1.2
V
2
2
J
.P. SELECTA s.a.u.
CODE 80342 REV H
9.2 List of Parameters
MIN
MAX
0
125
0
70
0
70
0
300
0
300
1
120
1
1.4
40
100
0
10
Note on the configuration parameters of the autoclave
operation.
1. If, accidentally, these parameters are modified, re-establish them
with the "default" value of this table.
2. Parameter B is important for adjusting the initial water volume of
the autoclave.
(For cycles with a long "purge" (more than 10 minutes) it is convenient
to increase this parameter about 5 ... 8s.
3. Parameter V establishes the difference between the water probe
and the "heart" probe.
The default value is 2ºC. Increasing this value to 4ºC, you may shorten
the treatment time due to a higher water temperature.
4. To access parameters, you must enter in Factory mode. Enter
password 0284 to establish this operator.
5. In addition to the previous parameters, there are 3 correction factors
for the different probes
Water probe: process control probe located at the bottom of the
chamber.
Product probe: probe placed inside the can for the control of the
product and the calculation of the F0.
Pressure probe: probe that controls the internal pressure of the
chamber.
Use a standard probe to adjust the autoclave reading values to the
value indicated by the pattern.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
JAN/2020
(It can be modified without notice)
UNITS
DESCRIPTION
1s
Filling maximum time
1s
Delay on detecting the water level
1s
-
1s
Maximum time water entry in cooling
1s
Vacuum time with pressure
x3min
Max. time to raise the temperature set
0.05bar
Absolute pressure to stop autoclave
1ºC
Open door temperature
Absolute pressure to stop the autoclave
1ºC
Offset heart probe
Page: 48