Check drain valve.
Check and clean the outlet filter.
Check the connection of the drain hose.
23
POWER FAIL Power cut off with the equipment operating
Cause:
During an operation cycle, the power has failed.
Resolution: Restart the cycle.
24
POWER FAIL RESUMABLE
Cause:
It indicates a cut off in the power while the equipment was operating.
The process can continue because the device has a battery and has preserved the status.
(If the temperature conditions during the cut have varied too much, another error may
appear, such as 36, as the chamber temperature has dropped too low).
29
SAFETY THERM
Cause:
Inner temperature has reached a too high value.
Resolution: • Wait for a few minutes and reset the thermostat
• Increase parameter "B" in 5s.
30
LEVEL ERROR
Cause:
No water enters the chamber.
Resolution: • Check the water inlet tap
• Increase parameter "B" so that more water enters.
34
INTERNAL UNDERTEMP Reading water probe temperature (connector plate:
K20) lower than 8ºC for more than 5 minutes during the HEAT phase or in any
moment during any other phase when the program is operating.
Cause:
• If this happens at the beginning of the cycle, it is likely that the water in the network is
very cold (winter).
Resolution
In the autoclave loaded with cans, pour half a bucket of hot water into the boiler and
then start the autoclave.
36
STERILIZ. UNDERTEMP Temperature drop
Cause:
In the sterilization phase, the temperature has dropped 2ºC below the set point for 30s.
Resolution: Check the temperature probe and the location inside the jar.
Make sure there has been no power interruptions.
37
STERILIZ. OVERTEMP Excessive temperature
Cause:
In the sterilization phase, the temperature has exceeded the set point at 4ºC for 5s.
38
EMPTY PRESSURE Timeout
Cause:
The temperature has been above the cooling temperature and the pressure has been
below the cooling pressure for 60 minutes.
39
STERILIZATION PRESSURE
Cause:
During the sterilization phase we have not reach the purge pressure in 30 seconds.
Resolution: Check that the back tap is not open, that the equipment is not leaking, that no solenoid
valve is open.
40
DISCONNECTED
Cause:
Communication with the power plate has been lost.
Resolution: Turn off the equipment and check connection between circuits.
41
PREHEAT timeout Excessive water pre-heating time
Cause:
The preheat temperature is not reached before 60'.
Resolution
• Check (Reset) the safety thermostat.
• Check the heating elements.
42
PURGE HEAT timeout The purge temperature is not reached before 60'
Resolution: • Check (Reset) the safety thermostat.
• The product is very cold, increase the temperature of the PRE phase.
J
.P. SELECTA s.a.u.
CODE 80342 REV H
Safety performance of the boiler thermostat
Water level is not detected
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
JAN/2020
(It can be modified without notice)
Page: 52
Tel 34 937 700 877 Fax 34 937 702 362

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

Steril-food 150l40024244002423

Tabla de contenido