8 Maintenance

8.1 List of failures

Error Code Faillure
E1-C1
Inner temperature probe error
E1-C2
Steam generator temp probe error
E1-C3
Outer temp probe error
E1-C4
Pressure sensor error
E2
Over pressure (>2.6bar)
E3
Chamber over temperature
E4
Keep warm
E5
Exhaust failed (20s< 0.3bar)
E6
Door open on working time
E7
Steam generator overtemperature
E8
Pre-heating fault >20min
E9
Filling steam gen. fails >30min
INCIDENCE: When the two TEST key leds are blinking means that a washing
cycle is necessary (every 100 cycles). With the door open, press SELECT
key (18) for 5 seconds. Leds B and D (20) will flash. Close the door and
press START.
Clean the water reservoir every week by using medical disinfectant. First drain the
distilled water in the reservoir, pumping the medical disinfectant and then wait for
an hour. Flush the reservoir with distilled water after medical disinfectant is drained.
Clean and disinfect the chamber every month by brushing the chamber inner surface
with medical alcohol.

8.2 Door adjusting

If you note that there are steam leakages through the door, you can adjust it by
doing the following:
1
Stop the autoclave and let it cold.
2
Insert the spanner into the gap beneath the plastic door cover.
3
Turn the door adjusting nut counter clockwise using the supplied spanner as
showed in the pictures.
NEVER ADJUST THE DOOR WHILE THE AUTOCLAVE
IS UNDER PRESSURE BECAUSE YOU MAY SUFFER
NEVER TRY TO ADJUST THE DOOR WHEN IT IS LOOKED.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
(It can be modified without notice)
Action to be taken
Check temperature probe
Check steam generator probe
Check outer probe temperature
Check pressure sensor
Check pressure sensor
Check temperature sensor
Keep pressure
Check valve
Check door look
Check temperature sensor
Check cylinder heaters
Check steam pipes
SCALDS AND BURNS.