Section 4: Precautions

4.1 System Precautions

• Equipment flushing and dental water supply system
maintenance are strongly recommended. See Section 10:
System Care.
• Close manual shut-off valve on the dental office water supply
every night before leaving the office.
• The use of an in-line water filter is recommended.
• Never operate system without fluid flowing through handpiece.
• Always ensure that the electrical connections on the handpiece
cable and the Steri-Mate® Handpiece are clean and dry before
assembling them for use.

4.2 Procedural Precautions

• The Cavitron SPS G90 unit works with Cavitron inserts as a system,
and was designed and tested to deliver maximum performance for all
currently available Cavitron and Cavitron Bellissima
inserts. Companies that manufacture, repair or modify inserts carry the
sole responsibility for proving the efficacy and performance of their
products when used as a part of this system. Users are cautioned to
understand the operating limits of their inserts before using in a clinical
setting.
• Like bristles of a toothbrush, ultrasonic insert tips "wear" with
use. Inserts with just 2 mm of wear lose about 50% of their scaling
efficiency. In general, it is recommended that ultrasonic inserts be
discarded and replaced after one year of use to maintain optimal
efficiency and avoid breakage. A DENTSPLY Professional Insert
®
brand ultrasonic
Efficiency Indicator is enclosed for your use
• If excessive wear is noted, or the insert has been bent, reshaped or
otherwise damaged, discard the insert immediately.
• Ultrasonic insert tips that have been bent, damaged, or reshaped are
susceptible to in-use breakage and should be discarded and replaced
immediately.
• Retract the lips, cheeks, and tongue to prevent contact with the insert tip
whenever it is placed in the patient's mouth.
• Water should be used for all scaling procedures. Water flow is adjustable
from less than 10 to greater than 60 ml per minute.
• Only adjust the systems's power adjustment knob with the insert outside
the patients mouth.
.