• Incoming water supply line pressure to the ultrasonic scaler
must be 25 psi (172kPa) minimum to 60 psi (414 kPa)
maximum. If your dental water system's supply line pressure
is above 60 psi, install a water pressure regulator on the
water supply line to your Ultrasonic Scaler.
• Incoming water temperature to the Cavitron System should
not exceed 25˚C (77˚F). If needed device should be installed
to maintain a temperature within this specification, or a
Cavitron DualSelect Dispensing System attached to allow
this system to be operated as a closed water system.
• A manual shut-off valve on the dental water system supply
line should be used so that the water can be completely shut-
off when the office is unoccupied.
• A filter in the dental water system supply line is
recommended so that the particles in the water supply will be
trapped before reaching the ultrasonic scaler.
• After the above installations are complete on the dental
water supply system, the dental office water line should
be thoroughly flushed prior to connection to the ultrasonic
system.
• After flushing system, verify there are no leaks.
ENGLISH • 8
Section 7: SPS™ Ultrasonic Scaler
System Description

7.1 System Controls

®
The DENTSPLY Cavitron
whenever the clinician removes the handpiece from the
handpiece holder.
ON/OFF Function
The Ultrasonics is switched ON and OFF using the
treatment unit's footswitch.
Power Adjustment
Power is adjusted by turning a potentiometer located on the
treatment unit head that has been installed by your service
technician. The potentiometer is provided in the Installation
Kit included with your unit.
Foot Controlled Power Boost Function
(optional)
A second position on the footswitch (when available)
provides the capability to briefly increase the units power
output for quick, efficient removal of tenacious deposits
using only the footswitch without touching the unit.
Lavage Flow Adjustment
Water flow through the handpiece is adjusted by rotating a
control element on the cable connector (blue).
SPS
module is enabled