Disposable Prophy Tip installed
in free air or without water for
an extended period as this may
cause the disposable prophy tip
to break.
• Do not use the SofTip
Disposable Prophy Tips with
any other ultrasonic insert or
sonic scaler tips. The Cavitron
SofTip Ultrasonic Implant Insert
is designed specifically for use
with these disposable tips and
use with any other ultrasonic
insert or sonic scaler tip may
cause the disposable tip to
break or dislodge.
• The Cavitron SofTip Ultrasonic
Implant Insert is designed for
use with Cavitron Ultrasonic
Scalers at low power setting.
6
Use at higher power settings
may cause the disposable
prophy tip to break.
• Should a disposable prophy tip
become dislodged and aspirated
by the patient, seek immediate
medical attention.
• Do not apply excessive pres-
sure when using the SofTip
Disposable Prophy Tip. A light
touch assures that the implant
fixture will maintain its structural
integrity and the implant site
will not experience excessive
vibrational trauma.
General Precautions
•
SofTip Disposable Prophy Tips
are intended for single use only.
Discard after each use.
Infection Control
Procedures
SofTip Disposable Prophy Tips are
intended for single use only.Discard
after each use.
Do not attempt to clean or sterilize
them. DO NOT REUSE THEM.
The Cavitron SofTip Ultrasonic
Implant Insert and the SofTip
Wrench should be sterilized after
each use following the Infection
Control Procedures outlined in the
Infection Control Procedures/
Directions for Use supplied with
each insert.
The metal SofTip Base used with
the SofTip Wrench should be
sterilized after each use following
the sterilization instructions
provided in the Maintenance
section of the Installation,
Maintenance and Operation
Instructions supplied with the
sonic scaler.
Cavitron Ultrasonic Scaler
Set-Up
Insert a Cavitron SofTip
Ultrasonic Implant Insert (C) into
a 30K Cavitron Ultrasonic Scaler
handpiece as you would any other
7