3.5.5 Finishing
1. Start finishing no earlier than 4  minutes  30  seconds after end of working time
(i.e. 6  minutes after activation) with cups, discs, and points (Enhance™ Finishing
System).
4

Hygiene

Cross-contamination.
Infection.
1. Do not reuse single use products. Dispose in accordance with local
2. Reprocess reusable products as described below.
4.1 Capsule Extruder 2

4.1.1 Manual cleaning

1. Right after use, remove capsule from the device and discard capsule according to
local regulations.
2. In case of gross contamination of the device, remove it using a soft brush and run-
ning water.
3. Discard gloves according to local regulations and disinfect hands with an appropri-
ate bactericidal, virucidal, and fungicidal hand disinfectant solution according to
local regulations. Use according to disinfectant solution manufacturer's Instructions
for Use.
4. Use a new clean pair of examination gloves.
5. Clean device with a soft brush, warm water, and an appropriate cleaning solution
until free of visible contamination. Pay special attention to device seams and inser-
tions.
6. Remove cleaning solution residue with a cloth soaked with water.
7. Dry device with a lint-free, single-use cloth.
Device can be manually cleaned for up to 1000 times.

4.1.2 Manual disinfection

1. After cleaning [4.1.1 Manual cleaning] thoroughly wipe all device surfaces with a
single-use cloth in combination with an alcohol-based, bactericidal, virucidal and
fungicidal disinfection solution approved according to local regulations and use
according to disinfectant solution manufacturer's Instructions for Use. Make sure
disinfectant solution is compatible with cleaning solution. Pay special attention to
device seams and insertions.
2. Remove disinfectant solution residue with a cloth soaked with water.
3. Dry device with a lint-free single-use cloth.
Device can be manually disinfected for up to 1000 times.
regulations.
7