4
Hygiene
4.1 Disposa-Shield
The use of the Disposa-Shield
ure against contamination, but does not substitute reprocessing of the SmartLite
curing light.
1. Remove Disposa-Shield
Cross-contamination.
Infection.
1. Disposa-Shield
2. Discard in accordance with local regulations.
4.2 Charger, handpiece, LED tip
Electrical short-circuit or dangerous malfunction.
Injury.
1. Disconnect the charger from the power source during reprocessing.
2. Safeguard charger, handpiece and LED tip against liquid penetration
4.2.1 Cleaning
1. Clean item using a moist cloth with water and detergent until the item is free of vis-
ible contamination.
4.2.2 Disinfection
1. Disinfect item using a moist cloth with an alcohol-based surface disinfectant (e.g.
Dürr FD 322, Ecolab Incidin Liquid, S&M Mikrozid AF Liquid, B. Braun Meliseptol
rapid) according to national/local regulations.
2. Follow the disinfectant manufacturer's Instructions for Use.
Items can be reprocessed for up to 3,600 times.
NOTICE
Wrong cleaning method.
Damage to SmartLite
1. Do not submerse into disinfection solutions.
2. Do not clean with highly aggressive disinfecting solutions (e.g. solu-
4.3 SmartLite
PS Eye Protection Shield
®
4.3.1 Cleaning
1. Clean item using a moist cloth with water and detergent until the item is free of vis-
ible contamination.
protective barrier
®
protective barrier is a mandatory precautionary meas-
®
protective barrier immediately after use.
®
protective barrier shields are intended for single use
®
only. Do not reuse in order to prevent cross-contamination or malfunc-
tion.
during cleaning and disinfection.
®
tions based on orange oil or acetone).
PS curing light.
PS
®