Table 1: 2D options
Control
Description
Adjusts the display brightness. Settings range from 1 to 10.
Brightness
The display brightness affects battery life. To conserve battery life, adjust brightness to a
lower setting.
Guide
Turns guidelines on and off. Guidelines are for needle guidance, are an optional feature,
and depend on transducer type.
For transducers with a single-angle or multi-angle bracket, the touchpad moves the
depth cursor.
If the transducer uses a multi-angle bracket, select Guide and then select the angle: A, B,
or C. To exit angle selection, select Back. To clear the guides, do either of the following:
Select the angle again (A, B, or C).
Exit angle selection and press Guide.
See also the needle guide's user documentation.
Sector
(Cardiac exam) Specifies the sector width.
SonoMB On is available only for Sector Full.
SonoMB
MB On and MB Off turn SonoMB™ multi-beam imaging technology on and off. When
(MB)
SonoMB is on, MB appears in the upper left-hand screen.
SonoMB depends on transducer and exam type.
ECG
Displays the ECG signal. See
This feature is optional and requires a FUJIFILM SonoSite ECG cable.
Clips
Displays the clips options. See
This feature is optional.
THI
Turns Tissue Harmonic Imaging on and off.
When on, THI appears in the upper left-hand screen. This feature depends on transducer
and exam type.
Page x/x
Indicates which page of options is displayed. Select to display the next page.
Notes
2D imaging
"ECG"
"To capture and save a clip"
If you use MBe frequently, use a shortcut key to turn on the MBe control.
For instructions to program a shortcut key, see
setup"
on page 1.
on page 4-27.
"A & B key and footswitch
on page 4-22.