NOTE: Curing Probes: The radiometer on the SmartLite IQ unit
charging base has been designed for use with the
13-8.5mm curing probe supplied with the unit only. Other
probes may lead to misleading readings and are not
recommended for use with the radiometer.

BATTERY LEVEL INDICATORS

LCD Display

The LCD displays information about the battery status in a battery
shaped icon. Black bars in the icon indicate the battery level
(see Figure 6). Bars are a relative indicator of battery strength.
3 and 4 Bars: No recharge required
2 Bars: Recharge recommended after completion of procedure
0 and 1 Bar: Recharge required prior to use

Battery Indicator Light

The battery indicator light (see Figure 4) on the top of the
handpiece above the LCD window provides information about the
battery level and charging status of the handpiece. The
information from this light also depends upon whether the
handpiece is on or off the charging base as described below:
• When the handpiece is removed from the charging base, the
battery indicator light indicates charge level showing a green,
amber, or red light corresponding to the level of charge of the
battery:
Green: No recharge required
Amber: Recharge recommended after completion of procedure
Red: Recharge required prior to use
NOTE: The SmartLite IQ unit has been designed to provide full
output power regardless of the charge status of the battery.
In the event that there is insufficient battery power to drive
the SmartLite IQ unit, the unit will automatically shut down,
three audible tones will sound, and the LCD will display
error code E1. See Errors.
• When the handpiece is placed onto the charging base, the
battery indicator light will be green, regardless of the state of
the battery.
• In the sleep state the battery indicator light turns off.

RECHARGING

The SmartLite IQ unit has been designed to provide sufficient
battery power for multiple days of use in a busy dental operatory
with a fully charged battery. For best results return the handpiece
to the charging base at the end of each day, or preferably after
each procedure.
1. Place the handpiece into the charging base. Be sure that the
bottom of the handpiece is properly seated so that the charging
pins of the charging base come into contact with the charging
pads of the handpiece (see Figure 4).
2. When the handpiece is properly seated in a powered charging
base, the battery indicator light on the handpiece will be green:
Blinking indicator light: Charging
Solid indicator light: Fully Charged
3. When the handpiece is properly seated in a powered charging
base, the LCD battery icon will display one of the following:
(see Figure 6).
Cycling Battery Icon: Charging
Solid Battery Icon: Fully Charged
4. When the handpiece is on the charging base, the curing time of
the LCD display is disabled. Simply remove the handpiece from
the charging base to restore these displays.

SLEEP STATE

The handpiece of the SmartLite IQ unit will enter a "Sleep" state to
minimize battery drain whenever it has been removed from its
charging base for more than 10 minutes without activation. It will
also enter this state 10 minutes after loss of power to the charging
base when the handpiece is on the charging base.
1. In Sleep State the LCD and battery indicator light on the top of
the handpiece will turn off.
2. To wake the unit from Sleep State and return it to Ready State,
it is necessary only to press any button on the handpiece.
3. Placing the handpiece on a properly powered charging base will
also wake the unit.
NOTE: The last used curing mode will be retained while the unit is
in its sleep state.
NOTE: The unit will not enter "sleep" state while in a properly
powered charging base.

ERRORS

The SmartLite IQ unit has been designed to operate with consis-
tent output power and for long life. In order to ensure the proper
operation of this light under normal use, several self-diagnostic
tests have been incorporated into the unit. These diagnostic tests
and the resulting error codes are described below. Note that any
error code is accompanied by a triple audible tone from the hand-
piece upon the start of the condition and whenever the handpiece
is triggered while in an error state.
In the event of a malfunction of the device, the LCD will show an
error code (e.g. "E1") in place of the curing time indicator
(Figure 9).
NOTE: The handpiece functions will fail to activate if any error
code is displayed in the LCD. See Error Codes below for
more information.
1. Error Code E1: This code results from an inability of the battery
to provide sufficient current to drive the SmartLite IQ unit.
Check to ensure that the handpiece battery is adequately
charged. The unit will automatically exit the error state upon
being recharged.
2. Error Code E2: This code results from improper contact
between the charging base and handpiece. Check to ensure
that the handpiece is seated properly on the charging base. The
unit will automatically exit the error state upon being properly
seated on the charging base.
3. Error Code E3: This code results from overheating of the
internal components of the light. It will be necessary to allow
time for the unit to cool before further use. The unit will
automatically exit the error state upon reaching operational
temperature.