Figure 2-12 : Installing the Ferrite Cores
Harsh Enclosure Cable Glands
The IND560 harsh environment terminal is designed to withstand severe
washdown environments. However, care must be taken when installing cables
and/or connectors that enter the terminal enclosure. To ensure a watertight seal:
•
Pass the cables through an appropriately sized cable grip before connecting
the wires. Figure 2-13 shows one load cell cable installed in its cable grip,
and a second grip disassembled.
•
Depending upon the diameter of the cable to be installed, select one of the two
different sized rubber grommets (if required) to properly seal around the cable.
Table 2-1: Grommet Cable Sizes
Grommet
None
Larger size hole
Smaller size hole
•
When making cable terminations inside the harsh enclosure, ensure that the
cable length from the terminal strip/connector to the terminal housing is
IND560 Installation Manual
Figure 2-13: Cable Glands
Cable Diameter
7–10 mm (0.28–0.39 in.)
5– 6 mm (0.20–0.24 in.)
3–4 mm (0.12–0.16 in.)
2-9