6.6
Changing the fresh air filter
NOTICE
i
i
With regular use and depending on the conditions of
use, the fresh air filter should be changed typically
every 2 months.
1. Unscrew the old fresh air filter.
2. Screw in the new fresh air filter.
6.7
Calibrating the touchscreen
1. At system start hold the function key until the calibration
screen is displayed.
2. Press each of the 5 position markers displayed in succession.
6.8
Cleaning
CAUTION
!
Impairment and wear of the system!
Heavy contamination or dust accumulation may impair
the function of the system and lead to increased wear
of individual system components.
Clean the system at regular intervals.
Abrasive
cleaning
cleaning agents and cleaning solvents can destroy the
fresh air filter.
The device does not need any special care.
If very dirty, the equipment can be carefully wiped down
with a damp cloth.
Carefully dry the device using a cloth.
7
Disposal
This product must not be disposed of as municipal
waste. This is indicated by the adjacent icon.
You can return this product to Dräger free of charge.
For information please contact the national marketing
organisations and Dräger.
Dräger X-dock 5300/6300/6600
implements
(brushes,
8
Technical data
Dimensions (H x W x D):
Master
Module
Weight:
Master
Module
Ambient conditions:
During operation
During storage
Gas connections:
etc.),
X-dock 5300/6300
X-dock 6600
Inlet pressure:
for the measured gas
for compressed air
Power supply:
Connections:
Serial no. (year of
manufacture):
CE mark:
Disposal
approx. 3.54 x 5.70 x 9.84 inches
(120 x 130 x 250 mm)
approx. 3.54 x 5.70 x 9.84 inches
(90 x 145 x 250 mm)
approx. 4.02 pounds
(48.23 ounces; 1500 g)
approx. 2.57 pounds
(30.86 ounces; 960 g)
32 °F to +104 °F (0 °C to +40 °C)
-4 °F to +122 °F (-20 °C
to +50 °C)
700 to 1300 hPa
max. 95 % relative humidity
1x fresh air connection
1x compressed air inlet
1x exhaust outlet
3x gas inlets
6x gas inlets
0.5 bar ±20 %
0.5 bar ±20 %
11 V - 28 V DC, 6.25 A
3x USB 2.0 standard A
connection, (host, cable <3 m)
1x USB 2.0 mini-B connection,
(device, cable <3 m)
1x Ethernet port RJ45
Data rate10/100 Mbit
The year of construction is given
by the 3rd letter in the factory
number located on the nameplate:
B=2010, C=2011, D=2012,
E=2013, F=2014, G=2015,
H=2016, etc.
Example: Serial number ARFH-
0054, the 3rd letter is F, so the
year of construction is 2014.
Electromagnetic compatibility
(Directive 2004/108/EC)
29