Basics
NOTICE
i
i
Dräger recommends the use of gases below the LEL
only. Gases above the LEL and below the UEL should
not be used.
Possible
safety
>100 %LEL include:
Using the maintenance station under a suitable
exhaust
Directly discharging the test gas into a suitable
exhaust or into the open air via an exhaust hose
(max. 10 m long).
Using pressure regulator valves with Flowstop
Opening the test gas cylinder only for the duration
of the test or calibration.
If the required expert knowledge is not available, it has to be
obtained from others (e. g. specialists, test institutions or
manufacturers).
NOTICE
i
i
Ensure adequate space for the entire assembly.
The master and all modules must have the same
firmware version. If this is not the case, a firmware
update needs to be carried out (see chapter 6.4 on
page 27).
1. If applicable, install modules on the master station in
accordance with the assembly instructions (X-dock 6300/
6600 only).
A maximum of 10 modules can be installed on one
master station.
The available modules can be combined in any way
desired.
2. If applicable, install wall mount or cylinder holder in
accordance with the assembly instructions.
3. Remove the sleeves from the intended gas inlets and from
the gas outlet.
NOTICE
i
i
If the sleeve is not removed from the gas outlet, the
station will be unable to conduct the self-test without
errors.
4. Fit the gas feed hoses to the gas inlets on the master and
connect them to the regulator valve on the test gas
cylinder.
NOTICE
i
i
For the gas exposure sequence Dräger recommends
using toxic gases with increasing concentrations.
Dräger recommends not to exceed a hose length of
10 m for the gas supply hoses.
5. If required, connect an exhaust hose (max. 10 m long) to
the exhaust outlet to discharge the test gas into an exhaust
or into the open air.
20
measures
when
using
6. Ensure a supply of compressed air or fresh air:
Connect the compressed air hose to the compressed
air port (outlet pressure of pressure regulator valve
0.5 bar, flow rate >3 L/min).
Adjust fresh air input (see chapter 4.6.1 on page 23).
Setting: Supplied by compressed air input
gases
OR
If required, connect the fresh air hose to the fresh air
filter.
Adjust fresh air input if necessary (see chapter 4.6.1 on
page 23). Setting: Supplied by pump.
WARNING
!
Risk of personal injury!
Impurities of the ambient air may result in faulty
measurements.
If you use the internal pump to supply fresh air through
the fresh air inlet you need to ensure that the ambient
air is devoid of interfering substances.
7. Connect the power pack.
Station with up to 3 modules: Power pack 24 V / 1.33 A
Station with 4 to 10 modules: Power pack 24 V / 6.25 A
The entire system is supplied with power via the master.
NOTICE
i
i
Dräger recommends the use of Dräger test gas
cylinders and Dräger pressure regulator valves (see
chapter 9 on page 30). Alternatively there is the option
of using a suitable pressure regulator valve with
0.5 bar outlet pressure and >3 L/min flow rate.
Dräger recommends to connect an exhaust hose
(max. 10 m long) to the exhaust outlet to discharge the
test gas into an exhaust or into the open air.
4
Basics
4.1
Switching the station on or off
NOTICE
i
i
If no action has occurred for 10 minutes, any user
logged in will be logged off automatically. After
45 minutes, the screensaver will be activated.
To switch the station on:
Press and hold the
1 second.
The following information is displayed during the switch-on
process:
Software version number
To switch the station off:
Press and hold the
3 seconds.
The station switches off.
key on the master for approx.
key on the master for approx.