Use
To set the gas cylinder level indicator:
NOTICE
i
i
The gas cylinder level indicator is only available for
cylinders that are configured via a Dräger part no.
1. Select Gas level monit.
2. Activate or deactivate check box Gas level monit.
3. Confirm selection with OK.
To reset the gas cylinder level indicator for a new test gas
cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select
> Station gas configuration.
3. Select desired gas inlet.
4. Select Further options and select
cylinder level indicator.
5
Use
WARNING
!
A defective pressure reducer on the gas cylinder can
lead to increased pressure in the station. The gas
hoses may loosen as a result and gas may escape.
Health hazard! Test gas must not be inhaled. Observe
the hazard warnings in the relevant Safety Data
Sheets.
Provide venting into a exhaust or
outside the building.
NOTICE
i
i
To prevent loss of gas, Dräger recommends closing
the gas cylinders when the station is left unattended for
long periods.
Adjustment may not be possible due to instrument and
channel errors.
When using a compressed air cylinder at the
compressed air inlet, Dräger recommends to always
equip all modules with gas monitors prior to starting
the test. Otherwise, the compressed air cylinder will be
emptied unusually quickly.
5.1
Conducting a visual inspection
A visual inspection of the instruments must be conducted
every time before being inserted into the station.
1. Check that the housing, external filters and the nameplates
are intact.
2. Check the battery contacts and sensor inputs for dirt.
NOTICE
i
i
Instruments that did not pass visual inspection must
not be inserted into the station. Otherwise the test
cannot be correctly assessed in its entirety.
24
5.2
to reset the gas
To insert the instrument in the module:
1. If necessary, push the lock up slightly and open the module
cover upwards.
2. Place the instrument in the corresponding module.
3. Close the module flap.
The instrument will be detected automatically.
X-am-125+ module with charging function only:
To remove the instrument from the module:
1. Push the lock up slightly and open the module cover
upwards.
2. Remove the instrument.
5.3
A self-test is carried out:
When the station is started up.
When the last successful self-test was more than 24 hours
ago and a test is being carried out.
The station is tested for leaks, pump function, and the software
version of the individual modules and the master.
Inserting or removing the gas measuring
device into or from a module
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
The charging function starts automatically approx.
15 minutes after the last test.
Station self-test
Dräger X-dock 5300/6300/6600
2
1
00233286.eps