Sensors
Pressure sensors:
Flow sensor:
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O (30 hPa) for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
Standard:
SlimLine or Standard and humidification:
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
Standard:
SlimLine or Standard and humidification:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Alarm volume settings
Low (nominal 54 dBA), Medium (nominal 60 dBA), High (nominal 73 dBA)
Physical - device and humidifier
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2015):
Weight (device and cleanable humidifier):
Housing construction:
Water capacity:
Cleanable humidifier - material:
Temperature
Maximum heater plate:
Cut-out:
Maximum gas temperature:
Air filter
Standard:
Hypoallergenic:
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
It is recommended that the device is a minimum distance of 0.8" (2 cm) from the body during operation. Not applicable to
masks, tubes or accessories. Technology may not be available in all regions.
Internally located at device outlet, analogueue gauge
pressure type, 0 to 40 cm H
Internally located at device inlet, digital mass flow type,
-70 to +180 L/min
O (40 hPa) for more than 1 sec.
2
25 dBA with uncertainty of 2 dBA
25 dBA with uncertainty of 2 dBA
27 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
35 dBA with uncertainty of 2 dBA
116 mm x 255 mm x 150 mm
22 mm
1336 g
Flame retardant engineering thermoplastic
To maximum fill line 380 mL
Injection moulded plastic, stainless steel and silicone seal
68°C
74°C
≤ 41°C
Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibres in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
2G GSM, 3G, 4G (LTE)
O (0 to 40 hPa)