Contact Abaxis Technical Support at 1-800-822-2947 to review the patients results printed on the error
report. You can print the error report using the Recall function starting from the Home screen. See
"Recalling Results" on page
Disclaimer:
Note that "~~~" is printed in the results for an individual chemistry when the
instrument iQC detects a possible fault condition in the chemistry reading or cal-
culation. If this happens, contact Abaxis Technical Support for error report inter-
pretation and further instructions on testing the sample.
8.3.2
Rotor Cancellation
In this situation:
1.
The display shows
2.
The display also shows an iQC code and a brief description of the problem.
Record the iQC code, and print an iQC code report using the Recall function, which can be accessed
from the Home screen. For details about printing these reports, see
8.3.3
Symbols on the Results Printout
These symbols print next to the result (see
The result is outside the reference range.
*
The result is lower than the dynamic range.
<
The result is higher than the dynamic range.
>
To obtain more information for results that have
(see
"Recalling Results" on page
approximation of the analyte concentration.
In addition, these symbols ("flags") are printed when a result is not available:
~~~
HEM
LIP
ICT
To obtain more information for results with (H) HEM, (L) LIP or (I) ICT, print an error report (see
"Recalling Results" on page
The entire numerical line displayed on the error report, including the line HEM, LIP, and ICT, must be
interpreted by an Abaxis Technical Support Representative to determine whether results are usable.
Call Abaxis Technical Support at 1-800-822-2947.
8-2
6-1.
and no patient results are printed.
Run Canceled
"Running Tests" on page
6-1). This report includes patient results, but the values are an
Chemistry interference
Hemolysis interference
Lipemia interference
Icterus interference
6-1). The error report shows a value with HEM, LIP or ICT next to it.
Troubleshooting
"Recalling Results" on page
or
printed before the value, print an error report
<
>
4-4):
6-1.