Using a Disposable 100 µL Capillary Pipette
1.
Keep the capillary pipette horizontal until it is over the sample
port.
2.
Raise the pipette to vertical (so it is perpendicular to the rotor sur-
face), and place its tip in the sample port, as shown at right.
3.
Push the black plunger down slowly and smoothly until the white
plug reaches the end of the capillary pipette. (There will be a slight
resistance at first.) See
delivery volume.
1.3.4
Step 4: Analyze the Sample
For the procedure, see
If problems occur, see
1-8
"Sample Delivery Volume" on page 1-7
"Running Tests" on page
"Troubleshooting" on page
Quick Reference Guide
4-4.
8-1.
for the appropriate sample