Sensor Settings Polytron ® 8Xx0; Setting Sensor Lock - Dräger Polytron 8000 Serie Instructions For Use

Ocultar thumbs Ver también para Polytron 8000 Serie:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

en
Sensor settings Polytron
If the build-up of deposits on the optical surfaces increases
beyond the point where a measurement is possible, the
instrument will go into fault.
● To set beam block signals on/off:
1. Select Settings > Communication > Analog interface
and the desired option.
Beamblock on/off
Set beamblock
2. Select Enable or Disable and confirm.
● To set beam block signals:
1. Select Settings > Communication > Analog interface
and the desired option.
Beambl. current
Beamblock limit.
2. Select the current line for editing the current and confirm.
3. Set the current and confirm.
 The setting for the beam block current is displayed.
4. Select Confirm and confirm with OK.
14
Sensor settings Polytron
14.1 Switching automatic calibration on or
off
1. Select Settings > Sensor > Auto calibration and
confirm.
2. Select Enable or Disable confirm with [OK].
14.2 Capture range
The capture range blanks out measurement fluctuations.
Measurement fluctuations are minor variations in measured
values (such as signal noise, variations in concentration).
Those variations do not change the transmitted or displayed
value. Measured values within this range will be displayed
with the capture value. Measured values outside the capture
range will be displayed with the actual measured value.
Capture value (offset)
The capture value is displayed continuously, as long as the
measured value ranges between the upper and lower capture
value limits.
Upper capture value limit
The upper capture value limit marks the upper limit of the
range in which the capture value is displayed.
®
8xx0
This function switches the
beam block on or off
This function sets the current
to the beam block warning
current
This function defines the cur-
rent for the beam block sig-
nal.
This function defines the
maximum limit if the beam
block signal will be transmit-
ted on the analog interface.
®
8xx0
Lower capture value limit
The lower capture value limit marks the lower limit of the
range in which the capture value is displayed.
14.2.1 Setting capture range
1. Select Settings > Sensor > Display Capture and
confirm.
2. Set offset in the first window.
3. Set the lowest value of the capture range and confirm with
Next.
4. Set the highest value of the capture range and confirm
with Next.
5. Check all the values and confirm with Next.
14.2.2 Switching capture range off
1. Select Settings > Sensor > Display Capture and
confirm.
2. Set offset in the first window.
3. Set the lowest value of the capture range to 0 and confirm
with Next.
4. Set the highest value of the capture range to 0 and
confirm with Next.
5. Check all the values and confirm with Next.
14.3 Resetting sensor
This function resets the sensor to factory default settings.
1. Select Settings > Sensor > Init. channel and confirm.
2. Select Confirm and confirm with [OK].
14.4 Setting calibration interval
After the calibration interval expires, the instrument will issue
a warning that the calibration is past due.
1. Select Settings > Sensor > Cal. interval and confirm.
2. Set the calibration interval and confirm with [OK].

14.5 Setting sensor lock

This point is only valid for Polytron
87x0 IR.
1. Select Settings >Sensor > Sensor lock and confirm.
2. Select On or Off.
On
The instrument rejects any sensor whose
part number does not match the part num-
ber of the previously installed sensor.
Off
The instrument accepts any suitable sen-
sor, uploading the default settings of this
sensor. However, this overwrites any cus-
tomized settings with the default settings of
the new sensor.
®
8100 EC and Polytron
®
Dräger Polytron
8000 Series
®