Error Messages; Error Conditions - Mettler Toledo ICS241 Instrucciones De Manejo

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Error
Cause
• Backlighting set too dark
Display dark
• No mains voltage
• Power supply not plugged in
• Brief fault
• Location with vibrations
Weight display
unstable
(restless installation location)
• Drift
• Contact between load plate
and/or weighing sample and
surroundings
• Incorrect zeroing
Incorrect weight
display
• Incorrect tare value
• Contact between load plate
and/or weighing sample and
surroundings
• Scale tilted
• Load plate not on the scale
• Weighing range not reached
(less than 9d)
• Weighing range exceeded
(more than 9d)
• Result not yet stable
44

Error messages

5 Error messages
5.1

Error conditions

Remedy
➜ Set backlighting brighter
➜ Check mains
➜ Plug in power supply
➜ Switch device off and on again
➜ Avoid vibration or adjut vibration filter
➜ Avoid drift
➜ Avoid contact between load plate and/or weighing sample and
➜ Unload scale, set to zero and repeat weighing operation
➜ Clear tare
➜ Avoid contact between load plate and/or weighing sample and
➜ Level scale
➜ Place load plate on the scale
➜ Set to zero
➜ Unload scale
➜ Reduce preload
➜ If necessary, adjust vibration filter
surroundings
surroundings
ICS241 30232251A
Tabla de contenido
loading

Tabla de contenido