Error messages

{.
|
.
|
.
|
.
|
.
}
<.
=
.
=
.
=
.
=
.
>
------
-.-.n.o.-.-
{.|.n.o.|.}
<.=.n.o.=.>
E.r.r.6.
N.
o
. . .
C
.
a
.
L
. . I . M . M . 8 .
E.r.r.5.3.
Err
low voltage
rESO?
*
corr
loco01ltage
ÑÖOre
*
corr
loco01ltage
Abort?
*
corr
loco01ltage
Overload
Unload the scale or reduce the preload.
Underload
Place the weighing pan on the scale and ensure that it can move freely.
Weighing result never becomes stable
1. Ensure the environment is stable
2. Ensure the weighing pan can move freely
3. Change the setting of the vibration adapter (Section 6.2)
Function not executed
The called function could not be executed.
Zero setting not possible (overload/underload)
Ensure that zeroing is not done with overload or underload.
No calibration/adjustment
Unplug power plug and plug in again. If message reappears, calibrate/adjust scale (Section
6.3). If this still does not help, contact your authorized METTLER TOLEDO representative.
EAROM checksum error
Unplug power plug and plug in again. If message reappears, contact your authorized
METTLER TOLEDO representative.
Undervoltage
The output voltage of the AC adapter is too low. Replace with an AC adapter with the correct
output voltage.
Readability
The resolution of the target weight which was entered when formula weighing is too fine.
Adjust to the readability of the scale (e.g. 15.01 g is entered with a readability of 0.05 g).
Target quantity not yet reached
The target quantity of the component has not yet been reached. Add more of the component
until the target quantity is reached.
Formula weighing terminated
Formula weighing has been terminated by pressing the «Last Comp» key and holding it
down.