Maintenance Menu - Mettler Toledo ICS241 Instrucciones De Manejo

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Test scale
Start?
Keyboard test
Start?
Display test
Start?
Serial number
Print setup
Reset all
Perform reset
30232251A ICS241

4.6 Maintenance menu

Test the scale.
Scales with an analog scale interface offer the following test procedure:
1. Press
to start the scale test.
The scale checks the zero point and -0- appears on the display.
The test weight value flashes in the display.
2. If necessary, adjust the test weight value displayed using
3. Put the test weight on the scale and confirm with
The scale checks the test weight.
After the test is complete, the deviation from the last calibration briefly appears on
the display, ideally *d=0.0 g.
The device switches to the next menu item.
Keyboard test.
1. Press
to start the keyboard test.
2. Press the keys in the order displayed on the screen.
3. If the key works, the device switches to the next key.
Display test.
1. Press
to start the display test.
A checkerboard pattern is displayed.
2. Press
to leave the display test.
The display works properly if the dark and bright fields are displayed without
missing pixels.
Display of the serial number of the counting scale.
Printout of a list of all menu settings.
Reset all settings to factory settings.
Confirmation inquiry.
• Reset all settings to factory settings with YES.
• Do not reset settings with NO.
.
.