Application
3.2.2 Automatic totalizing
Prerequisite
9 The function is activated in the Application –> Auto totalize menu.
9 The ∑ symbol is displayed in the symbols line.
9 Activate Good print for under/over checkweighing.
– or –
Activate Auto print for other weighing applications.
9 The
soft key is activated in the Terminal –> Device –> Keyboard –>
Soft keys menu.
9 If you want to display the totalized result on the screen, specify Total net in the auxiliary
line (Terminal –> Device –> Display –> Auxiliary line) and/or
accumulate number n for the Info key (Terminal –> Device –> Keyboard
–> Info key).
Automatic totalizing
1. Press
to enter the totalization screen in weighing or counting mode.
2. Load the first item.
If auto print is activated, automatic totalizing is executed and total net & number of
items are updated.
If the auxiliary line or info key is set, total net and number of items are displayed in
that area.
3. Unload the scale.
4. Load the next item.
The total net and number of items are increased.
5. Repeat steps 2 and 3 for further items.
If you need to cancel the former step, press
Note: This action only cancels the last operation.
6. When totalization is complete, press ∑.
A Save instruction is displayed.
7. Press YES to clear the totalized result.
– or –
➜ Press NO to continue totalizing.
Piece counting results and checkweighing/checkcounting results can be totalized the
same way.
.